User’s Manual 9R122-A
June, 2010
M203 Altimeter / Indicated Air Speed Tester
USER’S MANUAL
The Meriam Process Technologies (Meriam) M203 Altimeter / Indicated Air Speed Tester is a
microprocessor based pressure sensing device used to measure pressure 0-900 millimeters
(mm) Hg relative to absolute zero. The handheld is NIST traceable.
When used in Pressure Measure Mode, pressure can be displayed in a variety of engineering
units. Select the
ALT / IAS engineering unit and the M203 converts to Altitude Mode for checking
altimeter instruments. Pressing the TARE button from Altitude Mode changes the M203 to Air
Speed Mode for checking air speed instruments.
A timed leak test function is available for Pressure, Altitude and Air Speed Measure Modes. Other
feature
s
include a Min/Max capture, selectable damp rates and field re-calibration.
Page view 0
1 2 3 4 5 6 ... 31 32

Summary of Contents

Page 1 - USER’S MANUAL

User’s Manual 9R122-A June, 2010 M203 Altimeter / Indicated Air Speed Tester USER’S MANUAL The Meriam Process Technologies (Meriam) M203 Altimeter / I

Page 2

72. Press MIN/MAX and TARE keysat the same time. (See Figure 1 above. Top Line of display reads “ZEROING SOURCE:” Bottom line of display reads “REF TO

Page 3

8If an error is made use the back space ◄ key to move the cursor back to the incorrect digit. Press the UP ▲ or DOWN ▼ arrow keys to correct the value

Page 4 - June, 2010

9Units Select The standard units available on the M203 are: Inches of Mercury at 0° C (in Hg @ 0° C) (absolute) Millimeters of Mercury at 0° C (mm Hg

Page 5

10ALT/ IAS (Altitude / Indicated Air Speed) Mode The model M203 is capable of displaying altitude (feet or meters based on “U.S. Standard Atmosphere 1

Page 6

116. If METRIC units aredesired, press the ▲ or ▼ arrow key Top line reads “UNITS SELECT” and bottom line reads “METRIC”. Units are set to METERS and

Page 7

12Damp Rate Select Adjustable exponential type damping is available to steady the display when measuring pulsating pressures. The M203 has a range of

Page 8

13User Info Select (Accuracy, SW version, Mfr date, SN) The User Info Select registers are designed to provide the user with information on the hardwa

Page 9

14Auto Shut-Off Enabling the Auto Shut-Off feature allows the manometer to turn itself off after a user selected period of keypad inactivity. Selectab

Page 10

15Lockout Select Enabling the Lockout feature prevents unauthorized users from making changes to the configuration of the manometer. To enter the Pro

Page 11

16Header Name Follow the steps below to edit the Header Name. Keystroke Display1. From the Measure Modepress the PRGM key. Top line reads “PROGRAM MOD

Page 12

Safety Information Failure to follow all instructions could result in injury. Read, understand and follow all safety warnings and instructions provid

Page 13

17Contrast Select The Contrast Select register allows the user to adjust the character contrast of the LCD display to provide the best visibility for

Page 14

18Data Logging Data Logging can be used to record pressure measurements. Two record modes are supported: automatic and manual. In automatic mode, a pr

Page 15

19Leak Test The Leak Test feature allows the user to determine the leak rate in the pneumatic system being monitored. Once configured, Leak Test monit

Page 16

20Re-Calibration The Manometer can be re-calibrated in the field for zero, span, and linearity. The proper primary standards must be available prior t

Page 17

21RE-CALIBRATION – 1 Point EDIT and START To perform a 1-point re-calibration, apply a pressure between 50% and 100% of Full Scale and then follow the

Page 18

22RE-CALIBRATION – 5 Point EDIT To edit the calibration points for a 5 Point re-calibration follow the steps below. NOTE: If the factory default value

Page 19

23RE-CALIBRATION – 5 Point START To begin the 5-point re-calibration procedure, turn the unit OFF and follow the steps below. Keystroke Display1. Pres

Page 20

2411. Press the PRGM key atYES to save the Re-Calibration data or at NO to exit without saving. YES to save the Re-Calibration data or at NO to exit w

Page 21

25Specifications Type, Range and Display Resolution: 17.4-3 psia (900.00 mmHg) – XX.YYY (XXX.YY) Minimum / Maximum Display Values: Altitude: -2000.0 t

Page 22

Battery Operation: > 100 hours continuous use, 1 year shelf life, auto power off programmable at disabled, 10, 20, 30, 60 or 90 minutes Enclosure:

Page 23

Table of Contents Certification/Safety/Warnings ...1 User Interface ...

Page 24

battery polarity markings at the bottom of the compartment, as shown here. To install the four batteries: 1) Make sure polarity of battery matches th

Page 25

Service and Calibration If the M203 can not be zeroed, recalibrated or is damaged, it must be returned to the factory for servicing.

Page 26

For customer assistance please call your local Meriam representative or Meriam directly. Meriam Process Technologies 10920 Madison Avenue Cleveland, O

Page 27

User’s Manual 9R122-A June, 2010 1 Certification/Safety/Warnings Fire/Explosion Hazard. This instrument is not intrinsically safe. DO NOT use in a

Page 28

2User Interface 1.Keypad FunctionsON/OFF & BACK ◄ KEY Turns the manometer on and enters the unit into the Measure Mode. Pressing the key while in

Page 29

3PRGM & ENTER ► KEY Puts the manometer into the Program Mode from the Measure Mode. When in the Program Mode, pressing this key se

Page 30

42.Zeroing the ManometerThe M203 Precision Smart Manometer is a stable and precise instrument. However, on occasion the M203 should ha

Page 31

51.To zero the manometer using Referenced to Absolute Zero,start with the unit turned OFF and use the following keystroke sequence: Keystroke Display

Page 32

62.To zero the manometer using Factory Zero, start with the unitturned ON and in Measure Mode. Then use the following keystroke sequence: Keystroke Di