Meriam M4 Series Precision Calibrator Data Logger Manual User Manual Page 25

  • Download
  • Add to my manuals
  • Print
  • Page
    / 40
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 24
ZM4 Instruction Manual 9R87-B June 2012
25
Volts/mA Auto Step Setup
Figure 20: Volts/mA Auto Step Setup Display
Reference Function
Start Select to start the Auto Step simulation.
Mode: Volts/mA mode for the simulation. For
reference only, cannot be modified here.
Begin Value: Starting simulation value. Brings up edit
display to change the value.
End Value: Ending simulation value. Brings up edit
display to change the value.
Num Steps: Number of steps in the procedure. Brings up
edit display to adjust the value between 1
and 10.
Dwell Time: Time to pause at each step before
advancing to the next step. Brings up edit
menu to select the value between Manual, 5
or 10 seconds.
Startup Delay:
(not shown in Fig 20)
Delay after pressing Start key before
simulation begins. Brings up edit menu to
select the value between None, 30 or 60
seconds.
Page view 24
1 2 ... 20 21 22 23 24 25 26 27 28 29 30 ... 39 40

Comments to this Manuals

No comments